UM-2D

* 분해능 : 0.01mm
* 측정범위 : 1~300mm
3~18mm(코팅투과 모재 두께 측정시)
* 2 포인트 보정
* 저장기능:Up to 500 readings can be divided into a maximum of 5 files (user-selectable)

 제품특징

* 분해능 : 0.01mm

* 측정범위 : 1~300mm

                 3~18mm(코팅투과 모재 두께 측정시)

* 2 포인트 보정

* Data logger capacity: Up to 500 readings can be divided into a maximum of 5 files (user-selectable)

 

 제품사양

Operating Principle

Standard mode: P-E (pulse-echo) method with dual-crystal probes

Coating mode:  E-E (echo-echo) method with dual-crystal probes

Measuring range:

1.0mm to 300mm in standard mode

3.0mm to 18.0mm in through coating mode

Depending on probe, material and surface condition

Units and Resolution:

Mm - 0.01, 0.01

Inches - 0.001, 0.01

Probe Zero Adjustment:

Selectable: 1-point calibration: normal measurement

                   2-point calibration: curve surface measurement or other special application with a high precision

V-path correction:

Automatic

Measuring error:

Up to 10mm: ±0.05mm

Above 10mm: ±(0.5%H+0.01)mm

Repeatability:

±0.05mm

Display:

128×64 dot-matrix LCD screen with EL backlight (42mm×57mm)

Measurement update rate:

4 Hz

Material velocity range:

1000 -- 9999 m/s;

Data logger capacity:

Up to 500 readings can be divided into a maximum of 5 files (user-selectable)

Operating language:

English and Chinese

Operating time:

Up to 200 hours (without backlight) with alkaline batteries, depending on operating mode

      


   

 

 

 

 

 

 

 

 

Power:

Two 1.5V AA alkaline batteries (Warning with low battery voltage)

Auto shut-off:

After 5 minutes of non-use

Operating temperature:

-10℃ to +50℃; (Specification to -20℃ on request)

Size:

149mm×73mm×32mm (H×W×D)

 Weight:

200g; 7oz (including batteries)

 

Standard Package:

UM-2D thickness gauge

Standard probe (PT-08)

Carrying case

Two 1.5V AA alkaline batteries

Couplant

Operating manual

Certification of quality

Packaging list

Optional Accessories:

Attachable Rubber Sheath

Numerous special probes on request

Couplants and high-temperature couplant

4-step/5-step/6-step calibration block

Application software with Data transmission cable